The Elite electrical safety analyser features an array of powerful electrical safety tests including:
AC flash test or hipot test
DC flash test or hipot test
DC insulation resistance test
Earth bond test
Earth leakage test
Power / Run / Load test
Phase neutral continuity test
Analogue test
The Elite electrical safety tester is designed for automatic dual voltage (230V/115V, 50/60Hz) operation and is capable of testing product safety in accordance with most international safety standards and approvals agency recommendations – such as IEC, UL, CSA, VDE, EN, BEAB and BABT.
The Elite has a simple friendly user interface, centring on a single button operation with the option to allow engineers and supervisors to create their own pre-programmed test sequence for each product. Powerful Windows software is bundled with this product to enable the tester to be effectively network connected. The Elite 9024V incorporates a test capability where the function output test voltage can be varied from 10 to 280V.
The Elite not only carries out electrical safety tests but can also act as the core within an automated test system. It does this by acting as a simple PLC controller and can control up to 8 external I/O devices, reading inputs and making analogue sensor measurements. Using the sophisticated PC software, traceable test results can be stored and SPC type analysis can be undertaken on the measured data. The Elite incorporates a fast interlock system to ensure that operators are not exposed to hazardous conditions, and the output ports can be configured to drive beacons and other warning or safety shutdown devices. The Elite is the professional solution for practically every manufacturing or conformance electrical safety testing application, facilitating high levels of automation and process control within the production and design environment.
AC Hipot (flash) Test
Test voltage: 100V to 5kV AC in steps of 25V
Test voltage accuracy: ±2% ±20V
Test current: 20mA AC maximum
Measurement range: 0 to 19.99mA with a resolution of 0.01mA
Measurement accuracy: ±2% of reading ±2 digits ±20mA/kV
Ramp time: 0.5 to 25 seconds in steps of 0.1 second
Dwell time: 0 to 255 seconds in steps of 1 second
Failure modes: breakdown (dv/dt > 10V/ms)
software trip (user selectable 0 - 19.99mA)
hardware trip (22mA nominal)
Discharge time: 1 second maximum
Shut down time: <20ms maximum
DC Hipot (flash) Test
Test voltage: 200V to 6kV DC in steps of 25V
Test voltage accuracy: ±2% ±20V
Test current: 10mA DC maximum
Capacitance drive: 1mF
Measurement range: 0 to 9.99mA with a resolution of 0.01mA
Measurement accuracy: ±2% of reading ±2 digits
Ramp time: 0.1 to 25 seconds in steps of 0.1 second
Dwell time: 0 to 255 seconds in steps of 1 second
Failure modes: breakdown (dv/dt > 10V/ms)
software trip (user selectable 0 - 9.99mA)
hardware trip (11mA nominal)
Discharge time: 2 seconds maximum
Shut down time: <20ms maximum
Earth Continuity Test
Test voltage: 12V AC nominal into an open circuit
Test current: 0 to 26A in steps of 1A
Test current accuracy: ±3%
Measurement range: 0 to 1.999Ω with a resolution of 0.001Ω
Measurement accuracy: ±2% of reading ±2 digits ±5mΩ
Test time: 0 to 255 seconds in steps of 1 second
Protection: thermal overload
Measurement system: 4 wire
Insulation Resistance Test
Test voltage: 200 to 1050V DC in steps of 5V
Test voltage accuracy: ±2%
Break point current: 1mA DC minimum
Short circuit current: 1.5 mA DC nominal
Capacitance drive: 1mF at 10MΩ
Measurement range: 0 to 20MΩ in steps of 10kΩ, 19MΩ to 199.99MΩ in steps of 0.1MΩ
Measurement accuracy: ±2% of reading ±2 digits
Test time: 0 to 255 seconds in steps of 1 second
Phase Neutral Continuity Test
Test voltage: 9 Volts DC nominal into an open circuit
Test voltage accuracy: ±2%
Test current: 8mA DC short circuit
Measurement range: 0 to 1999Ω with a resolution of 1Ω
Measurement accuracy: ±2% of reading ±2 digits
Test time: 0 to 255 seconds in steps of 1 second
Protection: protection against connection to mains